roject name Specification parameters
Equipment Power Supply Specifications Equipment power AC380V 50Hz 4kw
Sample Power Supply Specification AC220V 16A
LWGYD-4 flow sensor Flow range :0.667-4.167 L/min, pulse output
LWGYD-15 flow sensor Flow range :10.00-100.00 L/min, pulse output
Big pumps 380VAC 50Hz 2.2kw protection grade: IP55 insulation grade: F
Mini Pumps 380VAC 50Hz 0.55kw protection grade: IP55 insulation grade: F
Scope of waterproof test

IPX1,IPX2,,IPX3,IPX4,,IPX4K,IPX4K,IPX5,IPX6,IPX6K,

IPX7,IPX8,JIS R1/R2,JIS S1/S2

IPX1-IPX2 Drop area :1000*1000 mm
  Drop hole diameter :0.4 mm
  Drop hole spacing :20 mm
  Test flow :1~1.5 mm/min (IPX1),3~3.5(IPX2)
IPX3-IPX4 Swing radius R400mm ,R600mm ,R800mm ,R1000mm
  Inner diameter of pendulum tube Φ18mm
  Pinhole diameter of pendulum tube Φ0.4mm, hole spacing 50 mm ,0.07±0.0035 L/Min per hole
  Swing Rainfall According to the test standard set on the console (water supply 0.63-10 L/Min)
  Test time Set on console according to test criteria (range 1-9999 seconds)
  Swing tube test angle 0-3500
  Swing tube test speed About 4-60 degrees per second
  Speed of swing tube turntable According to the standard, the pre-factory system has been set to 1 RPM, or 1 rpm
  Custom testing

Any set within the º of :0-350

Test speed :4-60 degrees per second set; unit: degrees per second

Test time :0-9999 seconds; set in seconds

Test flow rate :0.63-10.0 L/min set arbitrarily in L/min unit

Sprinkle water Number of fountains 121,φ0.5 holes
  Test flow 10±0.5 L/min
  Test time Set by test standard (range 1-9999 seconds) in seconds
  Test pressure 50-150 kPa
IPX5-IPX6 nozzle aperture 6.3 mm (IPX5),12.5 mm (IPX6)
  Water spray 12.5±0.625 L/min (IPX5),100±5 L/min (IPX6)
  Test time Set by test standard (range 1-9999 seconds) in seconds
  Test pressure Water flow regulation IPX5 about 30 kpa ,IPX6 about 100 kpa
IPX7 Test requirements

The lowest point of the shell with 1. height less than 850 mm shall be less than 1000 mm; above the water surface

The highest point of the enclosure 2. a height greater than or equal to 850 mm shall be 150 mm; below the surface of the water

  Test time 30 min
IPX4K Swing radius R200mm,R400mm,R600mm,R800mm
  Inner diameter of pendulum tube Φ22mm
  Pinhole diameter of pendulum tube Φ0.8mm 0.6 L/min±5 per cent per hole
  Angle of pendulum test 360°(±180° actual ±175)
  Test speed of pendulum tube s/Cycle 12
  Swing Rainfall 7.2-30 L/min±5% set
  Test time min 10(1 position 5 min,90°5)
  Water pressure 400 kPa
  Speed of turntable 1RPM
IPX6K Nozzle diameter Φ6.3mm
  Test flow 75L/min±5 per cent
  Test time 0-99999.9 s Set
  Spray pressure 1000 kPa
Turnover Speed range Tilt 15° RPM 1-20
  Adjust height mm 800-1700
Dimensions of Water Supply and Control System W1310mm×D890mm×H1600mm

 

latest company news about UL Guangzhou Laboratory Orders KINGPO IPX1-8 Rain Testing System and Digitization  0latest company news about UL Guangzhou Laboratory Orders KINGPO IPX1-8 Rain Testing System and Digitization  1latest company news about UL Guangzhou Laboratory Orders KINGPO IPX1-8 Rain Testing System and Digitization  2