logo
Send Message
products
PRODUCTS DETAILS
Home > Products >
KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing

KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing

Detail Information
Highlight:

ISO 27186:2020 Compliant High Voltage Current Carrying Test System

,

1.53 kV Peak Voltage Defibrillation Pulse Test System

,

50 A Peak Current ISO 27186:2020 Test Equipment

Product Description
KP-1240S High Voltage Current Carrying Test System

Technical Specification (Compliant with ISO 27186:2020 Annex E)

Product Overview

The KP-1240S High Voltage Current Carrying Test System is a specialized test instrument designed to generate high-intensity pulse currents simulating defibrillation shocks for active implantable medical devices. This system is designed and manufactured in accordance with the requirements of Annex E (normative) in ISO 27186:2020 — Active implantable medical devices — Four-pole connector system for implantable cardiac rhythm management devices — Dimensions and test requirements, serving as an updated supplement to the original Annex E in YY/T 0972-2016.

The primary purpose of this test system is to evaluate the four-pole connector system (IS-4/DF-4) in implantable cardiac rhythm management devices (such as pacemakers and ICDs) for permanent damage when subjected to external high-voltage defibrillation pulse currents, and to verify whether the original settings can be restored through reprogramming, thereby assessing the safety and reprogrammability of the connector.

Technical Specifications
Extended Performance Parameters
Parameter Specification
Voltage Range 0 - 2 kV (adjustable, with peak up to 1.53 kV as measured)
Current Peak ≥50 A (under standard 100 Ω load, simulating defibrillation conditions)
Pulse Duration 10 - 20 ms (adjustable; default 18 ms truncated exponential decay)
Waveform Rise Time < 10 µs (fast rise to simulate real defibrillator shocks)
Energy Delivery Up to approximately 200 J (calculated based on waveform and load)
Repetition Rate 1 - 10 pulses per minute (manual or automated sequencing)
Load Simulation Built-in resistive load (50 - 100 Ω) to mimic patient impedance during external defibrillation
Polarity Switching Automatic or manual selection of positive/negative polarity
Measurement Accuracy Voltage ±1%, Current ±2%, Time ±0.5%
Output Connections High-voltage safety connectors compatible with IS-4/DF-4 test fixtures
Safety and Compliance Features
  • Isolation: >10 kV isolation between output and control circuits
  • Discharge Mechanism: Automatic residual energy discharge after each pulse
  • Compliance Testing Modes: Pre-programmed sequences for multiple shocks (5-10 pulses as per standard protocols)
  • Environmental Specifications: Operating temperature 10-40°C, humidity <80% non-condensing
  • Power Supply: 220V AC ±10%, 50/60 Hz, <500W consumption
Output Waveform
  • Waveform Type: Monophasic truncated exponential waveform (simulating typical external defibrillator output)
  • Polarity: Positive or negative polarity selectable (meeting the standard requirements for monophasic waveforms)
Key Performance Parameters (Measured Waveform Data)
  • Maximum Peak Voltage: 1.53 kV (exceeding 1500 V, satisfying high-voltage shock requirements)
  • Peak Current in Test Circuit: ≥50 A (measured under standard load, ensuring sufficient shock intensity)
  • Pulse Duration: 18 ms (precisely controlled truncation time, compliant with exponential decay waveform characteristics)
Standard Compliance

This system is designed with ISO 27186:2020 Annex E as the core reference, which specifies the high-voltage pulse withstand test method for four-pole connector systems, including simulation of high-energy pulses applied to the connector by an external defibrillator.

Compared to YY/T 0972-2016, ISO 27186:2020 updates the dimensions, electrical, and functional requirements for four-pole connectors (IS-4 low-voltage type and DF-4 high-voltage type) and strengthens functional compatibility testing under high-voltage shocks.

The test pulse is designed to ensure no permanent damage to the connector after high-voltage impact and to support reprogramming recovery of the pulse generator, verifying overall system safety and reliability.

Measured Waveform Data
KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 0
Application Scenarios
  • Defibrillation shock withstand testing for four-pole connectors in implantable cardiac pacemakers
  • Safety evaluation of connector systems in ICDs (Implantable Cardioverter Defibrillators)
  • Compliance verification with the international standard ISO 27186:2020
  • Support for medical device registration and certification testing
Additional Features
  • High-precision waveform control: Digital adjustment of voltage, current, and duration for excellent repeatability
  • Safety protection mechanisms: Overload protection, short-circuit protection, and emergency stop function
  • Data recording: Built-in waveform acquisition and storage for easy test report generation
  • User interface: Touchscreen control with multi-language support (Chinese/English)
Why Choose the KP-1240S High Voltage Current Carrying Test System?

In today's increasingly globalized medical device industry, test equipment compliant with the latest ISO 27186:2020 standard has become essential for connector testing in implantable cardiac devices. The KP-1240S not only retains compatibility with YY/T 0972-2016 but has been fully upgraded to the 2020 international standard, helping companies efficiently meet CE marking, FDA approval, and other global registration requirements.

If you are conducting defibrillation withstand testing for implantable pacemaker or ICD connectors, the KP-1240S is your reliable choice, ensuring product safety and market competitiveness.

KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 1 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 2 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 3 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 4 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 5
Frequently Asked Questions
What standards does the KP-1240S comply with?
The system is fully compliant with ISO 27186:2020 Annex E for high-voltage pulse testing of four-pole connectors. It also maintains backward compatibility with YY/T 0972-2016.
What type of waveform does it generate?
It produces a monophasic truncated exponential waveform, accurately simulating external defibrillator shocks with a peak voltage of 1.53 kV, current ≥50 A, and 18 ms duration.
Is it suitable for both IS-4 and DF-4 connectors?
Yes, the KP-1240S supports testing for both low-voltage (IS-4) and high-voltage (DF-4) four-pole connector systems as defined in ISO 27186:2020.
How many pulses are applied during a standard test?
Typical protocols involve 5-10 pulses at maximum energy, but the system allows customizable sequences to match your specific test plan.
Does the system include test fixtures?
Basic high-voltage fixtures are included; custom adapters for IS-4/DF-4 connectors are available as options.
What safety features are built-in?
Overload protection, emergency stop, automatic discharge, and interlock mechanisms ensure operator safety during high-voltage operation.
Can it record and export test data?
Yes, built-in waveform capture and USB/PC interface for data logging and report generation.
What is the warranty and support?
Standard 1-year warranty, with optional extended service plans, calibration, and on-site training available.
How does this system help with regulatory approvals?
It provides traceable, repeatable results required for submissions to FDA, CE marking, and other global authorities.
Where can I get a demonstration or more details?
Contact our sales team for virtual demos, brochures, or quotes tailored to your testing needs.

Contact us for more technical details or a quotation!

Keywords: High Voltage Current Carrying Test System, ISO 27186:2020, Four-Pole Connector Testing, Implantable Cardiac Pacemaker Testing, Defibrillation Pulse Simulation, Medical Device Testing Equipment, DF-4 Connector Withstand Testing
products
PRODUCTS DETAILS
KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing
Detail Information
Highlight

ISO 27186:2020 Compliant High Voltage Current Carrying Test System

,

1.53 kV Peak Voltage Defibrillation Pulse Test System

,

50 A Peak Current ISO 27186:2020 Test Equipment

Product Description
KP-1240S High Voltage Current Carrying Test System

Technical Specification (Compliant with ISO 27186:2020 Annex E)

Product Overview

The KP-1240S High Voltage Current Carrying Test System is a specialized test instrument designed to generate high-intensity pulse currents simulating defibrillation shocks for active implantable medical devices. This system is designed and manufactured in accordance with the requirements of Annex E (normative) in ISO 27186:2020 — Active implantable medical devices — Four-pole connector system for implantable cardiac rhythm management devices — Dimensions and test requirements, serving as an updated supplement to the original Annex E in YY/T 0972-2016.

The primary purpose of this test system is to evaluate the four-pole connector system (IS-4/DF-4) in implantable cardiac rhythm management devices (such as pacemakers and ICDs) for permanent damage when subjected to external high-voltage defibrillation pulse currents, and to verify whether the original settings can be restored through reprogramming, thereby assessing the safety and reprogrammability of the connector.

Technical Specifications
Extended Performance Parameters
Parameter Specification
Voltage Range 0 - 2 kV (adjustable, with peak up to 1.53 kV as measured)
Current Peak ≥50 A (under standard 100 Ω load, simulating defibrillation conditions)
Pulse Duration 10 - 20 ms (adjustable; default 18 ms truncated exponential decay)
Waveform Rise Time < 10 µs (fast rise to simulate real defibrillator shocks)
Energy Delivery Up to approximately 200 J (calculated based on waveform and load)
Repetition Rate 1 - 10 pulses per minute (manual or automated sequencing)
Load Simulation Built-in resistive load (50 - 100 Ω) to mimic patient impedance during external defibrillation
Polarity Switching Automatic or manual selection of positive/negative polarity
Measurement Accuracy Voltage ±1%, Current ±2%, Time ±0.5%
Output Connections High-voltage safety connectors compatible with IS-4/DF-4 test fixtures
Safety and Compliance Features
  • Isolation: >10 kV isolation between output and control circuits
  • Discharge Mechanism: Automatic residual energy discharge after each pulse
  • Compliance Testing Modes: Pre-programmed sequences for multiple shocks (5-10 pulses as per standard protocols)
  • Environmental Specifications: Operating temperature 10-40°C, humidity <80% non-condensing
  • Power Supply: 220V AC ±10%, 50/60 Hz, <500W consumption
Output Waveform
  • Waveform Type: Monophasic truncated exponential waveform (simulating typical external defibrillator output)
  • Polarity: Positive or negative polarity selectable (meeting the standard requirements for monophasic waveforms)
Key Performance Parameters (Measured Waveform Data)
  • Maximum Peak Voltage: 1.53 kV (exceeding 1500 V, satisfying high-voltage shock requirements)
  • Peak Current in Test Circuit: ≥50 A (measured under standard load, ensuring sufficient shock intensity)
  • Pulse Duration: 18 ms (precisely controlled truncation time, compliant with exponential decay waveform characteristics)
Standard Compliance

This system is designed with ISO 27186:2020 Annex E as the core reference, which specifies the high-voltage pulse withstand test method for four-pole connector systems, including simulation of high-energy pulses applied to the connector by an external defibrillator.

Compared to YY/T 0972-2016, ISO 27186:2020 updates the dimensions, electrical, and functional requirements for four-pole connectors (IS-4 low-voltage type and DF-4 high-voltage type) and strengthens functional compatibility testing under high-voltage shocks.

The test pulse is designed to ensure no permanent damage to the connector after high-voltage impact and to support reprogramming recovery of the pulse generator, verifying overall system safety and reliability.

Measured Waveform Data
KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 0
Application Scenarios
  • Defibrillation shock withstand testing for four-pole connectors in implantable cardiac pacemakers
  • Safety evaluation of connector systems in ICDs (Implantable Cardioverter Defibrillators)
  • Compliance verification with the international standard ISO 27186:2020
  • Support for medical device registration and certification testing
Additional Features
  • High-precision waveform control: Digital adjustment of voltage, current, and duration for excellent repeatability
  • Safety protection mechanisms: Overload protection, short-circuit protection, and emergency stop function
  • Data recording: Built-in waveform acquisition and storage for easy test report generation
  • User interface: Touchscreen control with multi-language support (Chinese/English)
Why Choose the KP-1240S High Voltage Current Carrying Test System?

In today's increasingly globalized medical device industry, test equipment compliant with the latest ISO 27186:2020 standard has become essential for connector testing in implantable cardiac devices. The KP-1240S not only retains compatibility with YY/T 0972-2016 but has been fully upgraded to the 2020 international standard, helping companies efficiently meet CE marking, FDA approval, and other global registration requirements.

If you are conducting defibrillation withstand testing for implantable pacemaker or ICD connectors, the KP-1240S is your reliable choice, ensuring product safety and market competitiveness.

KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 1 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 2 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 3 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 4 KP-1240S High Voltage Current Carrying Test System with 1.53 kV Peak Voltage and ≥50 A Current Peak for ISO 27186:2020 Compliant Defibrillation Pulse Testing 5
Frequently Asked Questions
What standards does the KP-1240S comply with?
The system is fully compliant with ISO 27186:2020 Annex E for high-voltage pulse testing of four-pole connectors. It also maintains backward compatibility with YY/T 0972-2016.
What type of waveform does it generate?
It produces a monophasic truncated exponential waveform, accurately simulating external defibrillator shocks with a peak voltage of 1.53 kV, current ≥50 A, and 18 ms duration.
Is it suitable for both IS-4 and DF-4 connectors?
Yes, the KP-1240S supports testing for both low-voltage (IS-4) and high-voltage (DF-4) four-pole connector systems as defined in ISO 27186:2020.
How many pulses are applied during a standard test?
Typical protocols involve 5-10 pulses at maximum energy, but the system allows customizable sequences to match your specific test plan.
Does the system include test fixtures?
Basic high-voltage fixtures are included; custom adapters for IS-4/DF-4 connectors are available as options.
What safety features are built-in?
Overload protection, emergency stop, automatic discharge, and interlock mechanisms ensure operator safety during high-voltage operation.
Can it record and export test data?
Yes, built-in waveform capture and USB/PC interface for data logging and report generation.
What is the warranty and support?
Standard 1-year warranty, with optional extended service plans, calibration, and on-site training available.
How does this system help with regulatory approvals?
It provides traceable, repeatable results required for submissions to FDA, CE marking, and other global authorities.
Where can I get a demonstration or more details?
Contact our sales team for virtual demos, brochures, or quotes tailored to your testing needs.

Contact us for more technical details or a quotation!

Keywords: High Voltage Current Carrying Test System, ISO 27186:2020, Four-Pole Connector Testing, Implantable Cardiac Pacemaker Testing, Defibrillation Pulse Simulation, Medical Device Testing Equipment, DF-4 Connector Withstand Testing